HumaStar 600 ‐ Summary
Dr. M H B Ariyaratne edited this page May 16, 2025
·
1 revision
1. Communication Setup (Serial, ASTM E1381):
-
Configure RS-232 settings:
- Baud rate: 38400
- Parity: Odd
- Timeouts: 3000ms for query and retry, 100ms between messages
-
Ensure analyzer is set to Host Query Mode for automation:
- Auto send results = Enabled
- Auto request tests = Enabled
- Only order by LIS = Enabled
2. Implement ASTM Message Handling (ASTM E1394): You must handle the following message types:
- H – Header
- P – Patient Information
- O – Order
- R – Result
- Q – Query
- L – Terminator
Ensure your message parser and builder in ResponseHandler and LISCommunicator are structured to send and receive:
- Order queries (e.g., when a new barcode is scanned)
- Result transmissions (after result acceptance)
- Optional: Dilution Factor and QC results if needed
3. Sample Lifecycle in Host Query Mode:
- Analyzer sends a Q-record query for a sample.
- LIS replies with P + O records (patient + order).
- Analyzer runs test → sends back P + O + R + L messages (result).
- System must correctly respond to {ENQ}, {ACK}, {EOT}, {STX}, {ETX}, etc.
4. Message Construction Details:
-
^^^TESTIDformat required in O and R records. - Use consistent timestamp format:
YYYYMMDDHHMMSS. - Confirm that specimen descriptor values are mapped (Serum, Plasma, etc.).
- Use correct ASTM checksum calculation and block framing (important for hardware protocol).
5. Analyzer Software Settings:
-
Activate LIMS under:
Maintenance > Parameters > Software > LIMS. -
Enable settings:
- Auto send results
- Auto request tests
- Only order by LIMS
- Apply method limits
-
Leave QC/calibrator options disabled unless needed.
6. Testing & Troubleshooting:
- Use “LIMS > Communication” window on the analyzer to view live data flow.
- Use “LIMS > Save debug info” to access
ErrorsLog.txt. - Use “Purge” to reset LIS communications.
7. Optional Features (for future stages):
- Send Dilution Factor
- Query samples at rack positions
- Send individual result
- Handle batch and legacy modes (not recommended for new systems)