New CO2 stage #35
Comments
Moved original diode stage back to the diode |
configured x and y stages |
|
KP=0.185 these seem like good parameters. -0.003 immediately then takes a few seconds to completely eliminate the error |
Sign up for free
to join this conversation on GitHub.
Already have an account?
Sign in to comment
added new co2 stage with encoders. replaced X and Y. Z and the motion controller keep the same
needed to make new cables for X and Y. required limit signals to be inverted. added hex inverters to both cables.
updated stage1.ini and stage2.ini config files.
added EncoderType
changed DisplacementFullStep=0.0127 10TPI, 200 step == 0.0127 mm/step
set EncoderResolution=0.001 == 1 micron in mm