XR-5 Datalogger
SOFTWARE
Download software from here: http://www.pace-sci.com/LogXR.htm. Open LogXR
go to Port>Select>(select the appropriate COM port(COM4 for me))
if you don't know which COM port is for USB, open device manager to check
Setup>extract>
Setup should be as follows:
Log Mode: Manual/Slow
Log Interval: 10 seconds
Log Ambient: should be checked
Excitation Time: 30ms
Channel 1: OFF
Channel 2: ON
Description: Solar Radiation Sensor
Type: 0-2.5V
Mode: Standard
Slope: 600 (For units of W/m2)
Offset: 0
Channel 3: OFF
Channel 4: ON
Description: Sanguinipes temp
Type: Temperature PT9xx (C .rvt)
Mode: Average
Half-Range: selected
Slope: 1
Offset: 0
Channel 5: ON
Description: Sanguinipes temp
Type: Temperature PT9xx (C .rvt)
Mode: Average
Half-Range: selected
Slope: 1
Offset: 0
Channel 6: ON
Description: Soil temp
Type: Temperature PT9xx (C .rvt)
Mode: Average
Half-Range: selected
Slope: 1
Offset: 0
Channel 7: ON
Description: Ambient temp
Type: Temperature PT9xx (C .rvt)
Mode: Average
Half-Range: selected
Slope: 1
Offset: 0
Channel 8: ON
Description: Bivatatus temp
Type: Temperature PT9xx (C .rvt)
Mode: Average
Half-Range: selected
Slope: 1
Offset: 0
Channel X (Pulse tab): ON
Description: Avg wind speed
Type: Solid State Switch
Frequency: checked
Slope: 2.25 (wind speed in MPH – see page 14 for other units of measure)
Offset: 0
Start by selecting Logging>Start
Check to see if the green light on the data logger is blinking to make sure it is recording data
WIRE SETUP
Anemometer
Wire the WSD-100 Cable to the XR5 as follows:
Black X Wind speed contact closure
Green 1 Wind direction potentiometer
Yellow D 2.5v Potentiometer supply voltage
Red C Ground
Solar radiation
SRS-100 with 2 ft cable:
Green Channel 2 Output signal: 1.67 mV per W/m2
Red C Ground
Black C Ground
Yellow E Power: +5V DC, 1 mA
Temperature
PT9xx sensors:
Connect one lead of the temperature probe to the Channel #.
Connect the other lead to a ‘C’ terminal. Leads are not polarity sensitive.
Check that the sensors are working properly by selecting Real Time>Graph and looking at the readings.